Power supply:
Rechargeable battery and mains power
Temperature sensor:
Semiconductor 0-45ºC
Flow and colume transducer:
digital turbine
Flow / volume range:
± 16 Ls-1 / 10 L, BTPS
Volume accuracy:
±3% or 50 mL
Flow accuracy:
±5% or 200 mL/s
Dynamic resistance:
<0.8 cm H2O/L/s
Colour display:
FSTN 240 x 320
|
|
Printer /paper:
Thermal, 112 mm width
Mouthpieces:
30 mm external diameter
Serial output:
Rs 232, optoisolated
Dimensions:
310 x 205 x 65 mm
Weight:
1,9 Kg circa
Parameters:
FVC,
FEV1/FEV6%,PEF,FEF25-75%, FEF25%,FEF50%, FEF75%, FET, Vext, *FVC,
*FEV1, *PEF, PIF, PIVE,FIV1, *PEF, PIF, FIVC, FIV1, FIV1%, VC, IVC,
ERV, IC, VT, VE, BF, Ti, Te, Ti/Ttot, VT/Ti, MVV
Option:
carrying -case
|